USA Lab RE-10 Series Rotary Evaporator Board Programming Guide
Models: RE-1005, RE-1020, RE-1050. Summary: This guide explains how to program replacement control boards for RE-1005, RE-1020, and RE-1050 rotary evaporators. Replacement boards may come preconfigured for a different model, but they are fully interchangeable once the parameter values are updated.
Link to USA Lab RE-10 Series Rotary Evaporator Board Programming Guide
Quick Facts
- Applies To:
- R1005 – R1050
- Automatic 5L – Automatic 50L
- Required Action: Update parameters on the machine after board installation
Before You Begin
- If you have not yet installed the replacement board, refer to the:
- RE-10 Series Board Replacement Guide
STEP 1 — Power On the Rotary Evaporator
- Turn on the repaired rotary evaporator.
STEP 2 — Enter Parameter Menu
- Press and hold the silver knob until the parameter screen appears.
- Turn the knob until the display reads “68.”
- Press the knob to confirm.
STEP 3 — Update Parameters According to Your Model
- Use the table below to update each parameter:
|
Parameter |
RE-1005 |
RE-1020 |
RE-1050 |
|
Cc |
14 |
15.3 |
11.5 |
|
rP |
12 |
12 |
24 |
|
Pr |
5 |
5 |
5 |
|
2F |
0 |
0 |
1 |
|
rA |
4 |
4 |
4 |
|
rn |
120 |
120 |
120 |
|
P5 |
0000 |
0000 |
0000 |
- Turn the knob to move through parameters.
- Update each value to match your model.
STEP 4 — Save Settings
- On the final parameter (P5):
- Press the knob again to save all settings.
- The parameter menu will close automatically.
- The machine will return to the main menu.
- Congratulations — the control board is now fully programmed for your model.