USA Lab RE-10 Series Rotary Evaporator Board Programming Guide

Models: RE-1005, RE-1020, RE-1050. Summary: This guide explains how to program replacement control boards for RE-1005, RE-1020, and RE-1050 rotary evaporators. Replacement boards may come preconfigured for a different model, but they are fully interchangeable once the parameter values are updated.

Link to USA Lab RE-10 Series Rotary Evaporator Board Programming Guide

 


Quick Facts

  • Applies To:
    • R1005 – R1050
    • Automatic 5L – Automatic 50L
  • Required Action: Update parameters on the machine after board installation

Before You Begin

  • If you have not yet installed the replacement board, refer to the:
    • RE-10 Series Board Replacement Guide

STEP 1 — Power On the Rotary Evaporator

  • Turn on the repaired rotary evaporator.

STEP 2 — Enter Parameter Menu

  • Press and hold the silver knob until the parameter screen appears.
  • Turn the knob until the display reads “68.”
  • Press the knob to confirm.

STEP 3 — Update Parameters According to Your Model

  • Use the table below to update each parameter:

Parameter

RE-1005

RE-1020

RE-1050

Cc

14

15.3

11.5

rP

12

12

24

Pr

5

5

5

2F

0

0

1

rA

4

4

4

rn

120

120

120

P5

0000

0000

0000

  • Turn the knob to move through parameters.
  • Update each value to match your model.

STEP 4 — Save Settings

  • On the final parameter (P5):
    • Press the knob again to save all settings.
    • The parameter menu will close automatically.
    • The machine will return to the main menu.
  • Congratulations — the control board is now fully programmed for your model.