USA Lab RE-10 Series Rotary Evaporator Board Replacement Guide

This guide provides the full procedure for removing and replacing the circuit boards inside the RE-10 series control panel. Steps include removing the panel, disconnecting six connectors, removing wiring, taking out mounting screws, and reinstalling the new board.

Link to USA Lab RE-10 Series Rotary Evaporator Board Replacement Guide


Quick Facts

  • Boards Removed: 2
  • Total Connectors: 6 (3 on each board)
  • Total Mounting Screws: 8
  • Wires to Remove: RED, YELLOW, BLUE
  • Reinstallation: Reverse order

STEP 1 — Remove the Four Screws on the Back Panel

  • Remove the four screws circled on the back panel of the evaporator (pg.1).
  • Once removed, the panel can be opened to access the boards.

STEP 2 — Remove the Six Connectors

  • There are three connectors on each board.
  • Remove all six connectors circled in the images on page 1 (positions labeled 1–6).

STEP 3 — Remove the RED, YELLOW, and BLUE Wires

  • Disconnect the RED, YELLOW, and BLUE wires from the board.
  • Be sure to reinstall these wires in the same order on the new board.
  • (Image shown at top of page 2.)

STEP 4 — Remove the Eight Mounting Screws

  • Remove all eight screws holding the boards to the panel housing.
  • Once all screws are removed, the boards can be lifted out.
  • (Image shown on page 2.)

Reinstalling the Boards

Install the new boards in the reverse order of removal:

  1. Mount boards with the eight screws
  2. Reconnect the RED, YELLOW, BLUE wires in original order
  3. Reattach all six board connectors
  4. Reinstall the back panel